RESUMO
INTRODUCTION: The Woven EndoBridge (WEB) is a novel device for the treatment of wide-necked intracranial bifurcation aneurysms. The present series demonstrates our 'real-world experience' in the use of all iterations of WEB devices (available in Europe) in ruptured and unruptured aneurysms. METHODS: We analyzed our all-inclusive cerebrovascular database for patients treated with the WEB device between October 2010 and May 2015. Anatomic and clinical results are reported for all patients. RESULTS: One hundred and eight patients with 114 intracranial aneurysms were included in the series. Forty-seven aneurysms (41.2%) were ruptured. Eighty-six patients received angiographic and clinical follow-up after a mean of 13.4â months. One hundred and ten of 114 WEB devices (96.5%) were deployed successfully. Thromboembolic complications occurred in 11 of 110 interventions (10.0%), with a new permanent deficit in one patient. Re-rupture after WEB treatment was detected in two aneurysms (4.3%), which had both initially presented with subarachnoid hemorrhage. Angiographic follow-up revealed adequate occlusion in 68 of 90 aneurysms (75.6%). Fifteen aneurysms required retreatment. CONCLUSIONS: This series confirms a high level of safety and efficacy of the WEB device for the treatment of wide-necked intracranial aneurysms.